HT Wizard Easy-Touch Audiometer

Touch the future of hearing testing

Download Spec Sheet

Take your hearing testing program into the new millennium in one easy step. A touch-screen user interface and built-in OSHA compliance features are combined in the most effective automated hearing testing system ever developed -- the Tremetrics HT Wizard.

The Wizard audiometer features a large, easy-touch color display that conveniently adapts to any testing height and allows full-screen monitoring of real-time test data. To streamline data entry, customized subject questions and answers can be entered right on the screen. Operators enter data just once and the choices are then stored in convenient drop-down lists that activate with a touch. When it's time to test, the user simply touches the screen to start and spoken auto-test instructions are given in a choice of one of seven languages. This allows each subject to more confidently respond to the test tones, and assures a more reliable measure of their hearing acuity.

A host of time-saving features also help ensure OSHA compliance. The new patent-pending ID Wizard™ verifies and documents that all the earphone, handswitch, and OSCAR™ cables are correctly connected. Additionally, a daily Biological Log is automatically updated and can be printed out on demand. Comprehensive information on a particular audiometer and its calibration records are available at a touch of the display.

Up to 2,000 audiometric records can be stored, making the Wizard infinitely mobile and allowing comprehensive testing at any location. In addition, the audiometer will automatically calculate threshold shifts and PBI and provide instant archive-quality audiogram records via the quiet, built-in printer. For data transfer, an RS232C interface provides remote batch capability or serves as a direct link to mainframe computers or PC-based data management software such as Tremetrics FOSHM™.

Check the extra value features:

  • Large, adjustable-angle touch-entry display
  • Interactive multi-language subject instructions(U.S. Patent No. 5,811,681)
  • Separate left/right baseline reporting
  • Compliance documentation and Log
  • Headset auto-identification with patent-pending ID Wizard™
  • Extra-wide format, high speed printer
  • Storage for up to 2,000 audiogram records, backup for 2,000 additional records.
  • OSHA, Canadian and military (Military Entrance Processing Stations) categorization criteria.
  • Extensive management reports

HT-Wizard Audiometer Specifications

  • Equals or exceeds ANSI S3.6-1996.

  • Test frequencies: 500, 1000, 2000, 3000, 4000, 6000, 8000

  • Hz.Frequency accuracy: Better than 1% crystal controlled.

  • Distortion: Total harmonic distortion below 40 dB (1%)

  • HL attenuation: –10 dB to 100 dB in 5 dB steps.

  • HL accuracy: +/- 1 dB

  • Rise/Fall time: Meets ANSI specifications.

  • Calibration: Non-volatile EEPROM, calibration date printed on each audiogram. Calibration reminder 45 days prior to renewal. Provision for calibration of two headsets or headset/inserts.

  • Multimedia subject instruction languages: Currently seven - English, Spanish, Italian, French, German, Cambodian and Vietnamese.

  • Compliance documentation: Calibration; daily biological test; earphone, OSCAR and handswitch identification.

  • Headset: Telephonics TDH-39, 10 ohm earphones in Model MX41 cushions - includes automatic identification (ID Wizard™) for compliance reporting.

  • Handswitch: Long life, exceeding 1,000,000 closures - includes automatic identification (ID Wizard™) for configuration verification.

  • Insert earphone: EAR-Tone 5A

  • Mechanical: High impact, GE Noryl® U.L. approved plastic housing,

  • Display: 10.4" active matrix LCD color display with easy-touch user input. The display is adjustable over a 60-degree range so the screen can be set to the user's preference.

  • Physical: 17 1&Mac218;2" W x 13" D x 31&Mac218;2" H, weight 11.3 lbs.

  • Power: Auto-selectable 90 to 240 VAC with transient suppression for power line spikes.

  • Printer: High speed, extra-wide format graphics printer, 15 seconds with full compliance documentation.

  • Data output: High speed RS232C

  • Rear panel jacks: Four separate, standard 1/4" stereo jacks for left earphone, right earphone, OSCAR™ and handswitch.

  • Internal audiogram storage: 2,000

  • Internal audiogram backup: 2,000

  • Clock: Real-time, battery backed; time of day printout.