Here's how the Fit Tester 3000 works.

During a fit test, the respirator inlets are capped with test adapters, and the inhalation valves are removed from the mask. With the test subject holding his or her breath for no more than ten seconds, Fit Tester 3000 then establishes and maintains a slight vacuum, or controlled negative pressure, inside the mask. Since the respi-rator inlets are sealed, all sources of leakage into the mask are through the face-to-facepiece seal. The volume of air drawn out of the mask by FitTester 3000 during this short time period is equal to the leak rate into the mask through the face-to-facepiece seal. It's that simple...

Specifications

 

 

Dynamic Range

Leak Test Measurement

2 - 4500 cc/min

 

Fit Factor Computation

10 - 53,000

 

 

 

Pressure Sensors

Pressure Range

0 - 10 inches H20

 

Accuracy

+/- 0.5% full scale

 

Over-Pressure Limit

+ /- 5%

 

 

 

Instrument Accuracy

Challenge Pressure

+/- 5%

 

Leak Rate Measurement

+/- 3% or +/- 3 ml/min
(from 10 - 4500 cc/min)

 

 

 

General Specifications

Size

12.5" L x 5" H x 15" D
31.8 cm L x 12.7cm H x
38.1cm D

 

Weight

15 Ibs / 6.88 kg.

 

Power

110/220 VAC
(User Configurable)
50 Watts, 50/60 Hz.

 

 

 

Standard Accessories

Soft Vinyl Accessory Pouch;
Power Cord (Country Specific);
Squeeze Bulb Assembly Probe;
Cal Adapter User;
Dual Tube Assembly;
Manual

 